Intraradiological pathology-calibrated electrical impedance spectroscopy in the evaluation of excision-required breast lesions

Abstract

Purpose: Evaluating a real-time complementary bioelectrical diagnostic device based on electrical impedance spectroscopy (EIS) for improving breast imaging-reporting and data system (BI-RADS) scoring accuracy, especially in high-risk or borderline breast diseases. The primary purpose is to characterize breast tumors based on their dielectric properties. Early detection of high-risk lesions and increasing the accuracy of tumor sampling and pathological diagnosis are secondary objectives of the study.

Methods: The tumor detection probe (TDP) was first applied to the mouse model for electrical safety evaluations by electrical current measurement. Then it was utilized for characterization of 138 human palpable breast lesions that were to undergo core needle biopsy (CNB), vacuum-assisted biopsy (VAB), or fine needle aspiration (FNA) on the surgeon’s requests. Impedance phase slope (IPS) in frequency ranges of 100- 500 kHz and impedance magnitude in f = 1 kHz were extracted as the classification parameters. Consistency of radiological and pathological declarations for the excisional recommendation was then compared with the IPS values.